Customization: | Available |
---|---|
After-sales Service: | Yes |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
I.Introduction
The withstand voltage tester is an instrument for measuring the withstand voltage. It can intuitively, accurately, quickly and reliably test the electrical safety performance indicators such as breakdown voltage and leakage current of various objects to be tested, and can be used as a high voltage source to test the element. Device and overall performance. The pressure tester is designed in accordance with international and domestic safety standards such as IEC, CSA, UL, JIS, and withstand voltage output of 1kV~50kV and leakage current of 0~200mA. Suitable for all kinds of household appliances, power switches, wire and cable, transformers, terminal blocks, high-voltage bakelite electrical appliances, motors, medical, chemical, instrumentation, etc., as well as the test of safe withstand voltage and leakage current of high-voltage systems, and also a research laboratory And the pressure test equipment is indispensable for the technical supervision department.
The pressure tester product is based on the absorption and digestion of advanced pressure testers at home and abroad, combined with the actual use of many users to improve and perfect the design. The test voltage and leakage current are displayed at the same time. The function is rich and practical. It can reflect the actual value of the leakage current of the measured body through the leakage current display and compare the pressure resistance between different batches of different products or different manufacturers to ensure the safety of your product. The performance is flawless, and it is at the domestic leading level in terms of technical performance and quality reliability.
II.Sheet
MODEL |
2674AX |
||||||
A C W |
Output voltage |
(0.00~20.00)kV |
|||||
Max. output power |
200VA |
||||||
Max. current |
10mA |
||||||
Current range |
2mA,10mA |
||||||
Output waveform |
sine wave |
||||||
Waveform distortion |
≤5% |
||||||
Test time |
0.0s~999s 0.0= continue |
||||||
D C W |
Output voltage |
-------------------------------- |
(0.00~20.00)kV |
||||
Max. output power |
-------------------------------- |
200VA |
|||||
Max. current |
-------------------------------- |
10mA |
|||||
Current range |
-------------------------------- |
2mA,10mA |
|||||
Ripple factor |
≤5% |
||||||
Test time |
0.0s~999s 0= continue |
||||||
Volt meter |
Range |
(0.00~5.00)kV |
(0.00~20.00)kV |
||||
Accuracy |
±(5%+5 words ) |
||||||
resolution |
10V |
||||||
Displayed value |
RMS |
||||||
Ammeter |
Measure Range |
AC |
0.100mA ~ 200.00mA |
0.100mA ~ 10.00mA |
|||
DC |
----------------------------------- |
0.100mA ~ 10.00mA |
|||||
Resolution |
2mA:1uA, 20mA:10uA, 200mA:0.1mA |
||||||
Accuracy |
±(5%+5 words ) |
||||||
Timer |
Range |
0.0s~999s |
|||||
Min. resolution |
0.1s |
||||||
Accuracy |
±(1%+50ms) |
||||||
PLC interface |
no |
no |
Yes |
||||
Remote interface |
yes |