I.Instrument overview
This instrument is designed and manufactured according to the national standard GB/T8017 and the American national standard ASTM D323. It is suitable for measuring the vapor pressure of gasoline, volatile crude oil and other volatile petroleum products. It cannot be used to determine the vapor pressure of liquefied petroleum gas.Power supply: AC 220V±10%.
Instrument power consumption: ≤1100W
Experimental bomb: 3 samples at the same time, volume ratio of upper and lower chamber: 4:1.
Water bath temperature: 37.8°C.
Heating power: 1000W.
Stirring angle: 350 degrees.
Pressure gauge range: 0~0.16Mpa
III.Instrument installation
1. After unpacking the instrument, check whether the parts are complete according to the packing list (attached).
2. Check whether the surface of the instrument is intact and damaged. In particular, whether the LCD screen and the touch screen are well protected. If they are damaged, they cannot be powered on. Contact the manufacturer in time.
Overall diagram of the instrument
3. Before the instrument is turned on, check whether the connection of the 14-wire plug behind the electric control box is intact.
4. Add water to the bath, the water surface is consistent with the uppermost end of the gear.
5. With the power switch turned off, insert the power plug into the 220V mains.
6. Turn on the power switch, then the LCD screen should display the main interface, indicating that the instrument is turned on normally.